Post

Air-Powered Dental Handpieces and Motors - Performance Testing and Safety Requirements

This guidance covers air-powered dental handpieces and air motors (Class I reserved devices) with metal construction, maximum speed of 450,000 RPM, and intended for end-user sterilization. The devices are used in general dentistry for cutting/grinding teeth, cavity preparations, crown preparations, and finishing/trimming. The guidance excludes devices with non-metallic materials, electrical/battery components, prophy handpieces, bone cutting instruments, ultrasonic scalers, single-use devices, and devices not intended for user sterilization.

  1. Conduct comprehensive mechanical bench testing according to ISO 14457
  2. Perform usability evaluation following IEC 62366-1
  3. Complete biocompatibility testing for the three required endpoints
  4. Validate reprocessing procedures for 250 cycles
  5. Prepare detailed test reports and Declarations of Conformity
  6. Ensure clear labeling and instructions for use including reprocessing
  7. Verify compliance with all dimensional and performance specifications
  8. Document all test results in the format specified by FDA guidance
  9. Consider Pre-Submission if additional testing might be necessary
  10. Prepare comprehensive cleaning and sterilization validation documentation

Key Considerations

Non-clinical testing

  • Drop test according to ISO 14457
  • Noise level testing (should not exceed 80 dB)
  • Surface testing
  • Air supply testing (flow rate < 80 NL/min at 300 ± 100 kPa)
  • Water supply testing (if applicable)
  • Air and water pressure testing
  • Leakage testing
  • Speed testing (within ±10% of specified)
  • Eccentricity testing
  • Stall torque testing (if applicable)
  • Resistance to 250 reprocessing cycles

Human Factors

  • Usability evaluation according to IEC 62366-1

Labelling

  • Operating controls should use graphical symbols according to ISO 9687
  • Clear reprocessing instructions including recommended cleaning agents

Biocompatibility

  • Cytotoxicity testing
  • Sensitization testing
  • Oral Mucous Irritation testing

Safety

  • Operating controls design to minimize accidental activation
  • Connection safety requirements according to ISO 9168 and ISO 3964
  • Chuck system safety requirements for holding rotary instruments

Other considerations

  • ISO 14457: Dentistry – Handpieces and motors
  • ISO 9687: Dentistry – Graphical symbols for dental equipment
  • IEC 62366-1: Medical devices — Application of usability engineering to medical devices
  • ISO 9168: Dentistry — Hose connectors for air driven dental handpieces
  • ISO 3964: Dentistry – Coupling dimensions for handpiece connectors
  • ISO 10993-1: Biological evaluation of medical devices – Part 1: Evaluation and testing within a risk management process
  • ISO 7405: Dentistry – Evaluation of biocompatibility of medical devices used in dentistry

Original guidance

  • Air-Powered Dental Handpieces and Motors - Performance Testing and Safety Requirements
  • HTML / PDF
  • Issue date: 2024-09-30
  • Last changed date: 2024-09-27
  • Status: FINAL
  • Official FDA topics: Medical Devices, 510(k), Dental, Premarket
  • ReguVirta summary file ID: a238b3c3a80e24c5f68bd4d1a3195571
This post is licensed under CC BY 4.0 by the author.